Legato® 100 Syringe Pump

Legato® 100 Syringe Pump
Single Syringe Infusion Pump
Order Code: 788100

Infuse only single syringe pump. Accommodates syringes from 0.5 µl to 60 ml. User definable flow rates with selectable target volume or time values to control the total infusion volume.
  • Holds one syringe from 0.5 µl to 60 ml
  • Stall detection
  • High resolution color touch screen
  • Unparalleled ease of use
  • Password protected touch display
  • Run LED light on front panel
  • Full metal chassis
  • Built in syringe table
  • Up to 30 lb/13kg linear force
  • Advanced microstepping techniques
  • Built in RS485 interface
  • USB port
  • I/O & TTL interface
  • Continuous mode of operation
  • Spill dam
Accuracy
  • ±0.5%
Syringes (Min/Max)
  • 0.5 µl to 60ml
Flow Rate
  • Minimum (0.5 µl syringe): 1.28 pl/min
  • Maximum (10ml syringe): 25.99 ml/min
  • Maximum (60ml syringe): 88.28 ml/min
Display
  • 4.3 WQVGA TFT Color Display with Touchpad
Non-Volatile Memory
  • Stores all Settings
Connectors
  • RS485 - IEEE-1394 6 pos
  • USB - Type B
  • I/O & TTL - 15 pin D-Sub Connector
Linear Force (Max)
  • 13.6 kg (30 lb) @ 100% Force Selection
Drive Motor
  • 0.9 Stepper Motor
Motor Drive Control
  • Microprocessor with 1/16 microstepping
Number of Microsteps
per one rev.of Lead Screw
  • 15,360
Step Revolution
  • 0.069 µm/step
Step Rate - Minimum
  • 27.5 sec/µstep
Step Rate - Maximum
  • 26 µsec/µstep
Pusher Travel Rate - Minimum
  • 0.15 µm/min
Pusher Travel Rate - Maximum
  • 159 mm/min
Power
  • 100-240 VAC: 50/60 Hz, 50 W, 0.5 A fuse
Dimensions
  • 22.6 x 19.05 x 15 cm (9 x 7.5 x 5 in)
Weight
  • 2.66 kg (5.9 lb)
Operating Temperature
  • 4 degree C to 40 degree C (40 degree F to 104 degree F)
Storage Temperature
  • -10 degree C to 70 degree C (14 degree F to 158 degree F)
Humidity
  • 20% to 80% RH non condensing
Mode of Operation
  • Continuous
Classification
  • Class 1
Pollution Degree
  • 1
Installation Category
  • II
Regulatory Certification
  • CE, UL, CSA, CB Scheme, EU RoHS